Author (up) Vijande, J.; Carmona, V.; Lliso, F.; Ballester, F.; Perez-Calatayud, J. doi  openurl
  Title An efficient component of the redundancy calibration program to ensure equipment stability by assaying HDR Ir-192 sources at the time of replacement Type Journal Article
  Year 2024 Publication Journal of Applied Clinical Medical Physics Abbreviated Journal J. Appl. Clin. Med. Phys  
  Volume Issue Pages 5pp  
  Keywords brachytherapy  
  Abstract BackgroundBrachytherapy (BT) treatments involving temporary high-dose rate (HDR) sources are extensively employed in clinical practice. Ensuring the consistency of all measurement equipment at the hospital level is crucial, requiring a robust redundancy and consistency program. This enables the institution to verify the stability of the dosimetry system over time.PurposeTo describe, justify, and analyze a component of the redundancy program of the calibration protocols followed by the Radiotherapy Department of the Hospital Universitari i Polit & egrave;cnic La Fe (Val & egrave;ncia, Spain) during the last 10 years for the case of HDR BT as an additional component to ensure long term stability of the measurement equipment.MethodsAt the time the HDR BT source is replaced, its Air Kerma Strength (SK) is measured. By comparing this value with the one obtained at the time of installation (corrected by decay), a clear determination of the stability of the measurement equipment can be performed.ResultsDifference between SK,vendor and SK,hosp as a function of the measurement date is reported for a 10 years' period. All measurements are well within the +/- 3% tolerance level recommended in current international guidelines. Percentage differences of SK,hosp values at the time of replacement compared to SK,hosp ones at the time when the source was installed are within the +/- 0.5% range, reflecting oscillations around a null deviation.ConclusionsThe method proposed allows any hospital to ensure a redundancy component of the long-term stability of all equipment involved in BT measurements in a very simple and time efficient manner. Additionally, it enables the hospital to maintain a detailed log of historical differences, facilitating the identification and correction of potential systematic deviations over time.  
  Address [Vijande, Javier; Ballester, Facundo] Univ Valencia UV, Dept Fis Atom Mol & Nucl, Burjassot, Spain, Email: Javier.Vijande@uv.es  
  Corporate Author Thesis  
  Publisher Wiley Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 1526-9914 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:001306138000001 Approved no  
  Is ISI yes International Collaboration no  
  Call Number IFIC @ pastor @ Serial 6250  
Permanent link to this record
Select All    Deselect All
 |   | 
Details
   print

Save Citations:
Export Records:
ific federMinisterio de Ciencia e InnovaciĆ³nAgencia Estatal de Investigaciongva