|
|
Fernandes, L. M. P., Freitas, E. D. C., Ball, M., Gomez-Cadenas, J. J., Monteiro, C. M. B., Yahlali, N., et al. (2010). Primary and secondary scintillation measurements in a Xenon Gas Proportional Scintillation Counter. J. Instrum., 5, P09006–15pp.
Abstract: NEXT is a new experiment to search for neutrinoless double beta decay using a 100 kg radio-pure high-pressure gaseous xenon TPC. The detector requires excellent energy resolution, which can be achieved in a Xe TPC with electroluminescence readout. Hamamatsu R8520-06SEL photomultipliers are good candidates for the scintillation readout. The performance of this photomultiplier, used as VUV photosensor in a gas proportional scintillation counter, was investigated. Initial results for the detection of primary and secondary scintillation produced as a result of the interaction of 5.9 keV X-rays in gaseous xenon, at room temperature and at pressures up to 3 bar, are presented. An energy resolution of 8.0% was obtained for secondary scintillation produced by 5.9 keV X-rays. No significant variation of the primary scintillation was observed for different pressures (1, 2 and 3 bar) and for electric fields up to 0.8 V cm(-1) torr(-1) in the drift region, demonstrating negligible recombination luminescence. A primary scintillation yield of 81 +/- 7 photons was obtained for 5.9 keV X-rays, corresponding to a mean energy of 72 +/- 6 eV to produce a primary scintillation photon in xenon.
|
|
|
KM3NeT Collaboration(Aiello, S. et al), Alves Garre, S., Bariego-Quintana, A., Calvo, D., Cecchini, V., Garcia Soto, A., et al. (2025). Evaluation of the upgraded 3-inch Hamamatsu photomultiplier for the KM3NeT Neutrino Telescope. J. Instrum., 20(7), P07054–15pp.
Abstract: The 3-inch Hamamatsu R14374-02 photomultiplier tube is an improved version of the R12199-02 model and its successor in the construction of the KM3NeT neutrino telescope. A total of 1000 photomultipliers were analysed to assess their dark count rate, transit time spread, and spurious pulses. A subset of 200 photomultipliers were further evaluated to determine their quantum efficiency which is an essential parameter for Monte Carlo simulations of the detector response. The measurements show that R14374-02 model has better quantum efficiency homogeneity over the photocatode and better time properties than the R12199-02.
|