Llosa, G., Barrillon, P., Barrio, J., Bisogni, M. G., Cabello, J., Del Guerra, A., et al. (2013). High performance detector head for PET and PET/MR with continuous crystals and SiPMs. Nucl. Instrum. Methods Phys. Res. A, 702, 3–5.
Abstract: A high resolution PET detector head for small animal PET applications has been developed. The detector is composed of a 12 mm x 12 mm continuous LYSO crystal coupled to a 64-channel monolithic SiPM matrix from FBK-irst. Crystal thicknesses of 5 mm and 10 mm have been tested, both yielding an intrinsic spatial resolution around 0.7 mm FWHM with a position determination algorithm that can also provide depth-of-interaction information. The detectors have been tested in a rotating system that makes it possible to acquire tomographic data and reconstruct images of Na-22 sources. An image reconstruction method specifically adapted for continuous crystals has been employed. The Full Width at Half Maximum measured from a point source reconstructed with ML-EM was 0.7 mm with the 5 mm crystal and 0.8 mm with the 10 mm crystal.
|
|