Jordan, D., Tain, J. L., Algora, A., Agramunt, J., Domingo-Pardo, C., Gomez-Hornillos, M. B., et al. (2013). Measurement of the neutron background at the Canfranc Underground Laboratory LSC. Astropart Phys., 42, 1–6.
Abstract: The energy distribution of the neutron background was measured for the first time at Hall A of the Canfranc Underground Laboratory. For this purpose we used a novel approach based on the combination of the information obtained with six large high-pressure He-3 proportional counters embedded in individual polyethylene blocks of different size. In this way not only the integral value but also the flux distribution as a function of neutron energy was determined in the range from 1 eV to 10 MeV. This information is of importance because different underground experiments show different neutron background energy dependence. The high sensitivity of the setup allowed to measure a neutron flux level which is about four orders of magnitude smaller that the neutron background at sea level. The integral value obtained is Phi(Hall A) = (3.44 +/- 0.35) x 10(-6) cm(-2) s(-1).
|