Author Torres-Espallardo, I.; Diblen, F.; Rohling, H.; Solevi, P.; Gillam, J.; Watts, D.; Espana, S.; Vandenberghe, S.; Fiedler, F.; Rafecas, M. doi  openurl
  Title Evaluation of resistive-plate-chamber-based TOF-PET applied to in-beam particle therapy monitoring Type Journal Article
  Year (down) 2015 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 60 Issue 9 Pages N187-N208  
  Keywords PET; in-beam; RPC; particle therapy; TOF; range deviation; partial-ring  
  Abstract Particle therapy is a highly conformal radiotherapy technique which reduces the dose deposited to the surrounding normal tissues. In order to fully exploit its advantages, treatment monitoring is necessary to minimize uncertainties related to the dose delivery. Up to now, the only clinically feasible technique for the monitoring of therapeutic irradiation with particle beams is Positron Emission Tomography (PET). In this work we have compared a Resistive Plate Chamber (RPC)-based PET scanner with a scintillation-crystal-based PET scanner for this application. In general, the main advantages of the RPC-PET system are its excellent timing resolution, low cost, and the possibility of building large area systems. We simulated a partial-ring scannerbeam monitoring, which has an intrinsically low positron yield compared to diagnostic PET. In addition, for in-beam PET there is a further data loss due to the partial ring configuration. In order to improve the performance of the RPC-based scanner, an improved version of the RPC detector (modifying the thickness of the gas and glass layers), providing a larger sensitivity, has been simulated and compared with an axially extended version of the crystal-based device. The improved version of the RPC shows better performance than the prototype, but the extended version of the crystal-based PET outperforms all other options. based on an RPC prototype under construction within the Fondazione per Adroterapia Oncologica (TERA). For comparison with the crystal-based PET scanner we have chosen the geometry of a commercially available PET scanner, the Philips Gemini TF. The coincidence time resolution used in the simulations takes into account the current achievable values as well as expected improvements of both technologies. Several scenarios (including patient data) have been simulated to evaluate the performance of different scanners. Initial results have shown that the low sensitivity of the RPC hampers its application to hadron  
  Address [Torres-Espallardo, I.; Solevi, P.; Gillam, J.; Rafecas, M.] UV, CSIC, Inst Fis Corpuscular IFIC, E-46071 Valencia, Spain, Email: irene.torres@uv.es  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000354104700003 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 2227  
Permanent link to this record
 

 
Author Hueso-Gonzalez, F.; Vijande, J.; Ballester, F.; Perez-Calatayud, J.; Siebert, F.A. doi  openurl
  Title A simple analytical method for heterogeneity corrections in low dose rate prostate brachytherapy Type Journal Article
  Year (down) 2015 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 60 Issue 14 Pages 5455-5469  
  Keywords brachytherapy; low dose rate; heterogeneities; prostate; calcifications  
  Abstract In low energy brachytherapy, the presence of tissue heterogeneities contributes significantly to the discrepancies observed between treatment plan and delivered dose. In this work, we present a simplified analytical dose calculation algorithm for heterogeneous tissue. We compare it with Monte Carlo computations and assess its suitability for integration in clinical treatment planning systems. The algorithm, named as RayStretch, is based on the classic equivalent path length method and TG-43 reference data. Analytical and Monte Carlo dose calculations using Penelope2008 are compared for a benchmark case: a prostate patient with calcifications. The results show a remarkable agreement between simulation and algorithm, the latter having, in addition, a high calculation speed. The proposed analytical model is compatible with clinical real-time treatment planning systems based on TG-43 consensus datasets for improving dose calculation and treatment quality in heterogeneous tissue. Moreover, the algorithm is applicable for any type of heterogeneities.  
  Address [Hueso-Gonzalez, Fernando; Vijande, Javier; Ballester, Facundo] Univ Valencia, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain, Email: Javier.Vijande@uv.es  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000357620400009 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 2296  
Permanent link to this record
 

 
Author Solevi, P.; Magrin, G.; Moro, D.; Mayer, R. doi  openurl
  Title Monte Carlo study of microdosimetric diamond detectors Type Journal Article
  Year (down) 2015 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 60 Issue 18 Pages 7069-7083  
  Keywords ion-beam therapy; microdosimetry; diamonds; GATE  
  Abstract Ion-beam therapy provides a high dose conformity and increased radiobiological effectiveness with respect to conventional radiation-therapy. Strict constraints on the maximum uncertainty on the biological weighted dose and consequently on the biological weighting factor require the determination of the radiation quality, defined as the types and energy spectra of the radiation at a specific point. However the experimental determination of radiation quality, in particular for an internal target, is not simple and the features of ion interactions and treatment delivery require dedicated and optimized detectors. Recently chemical vapor deposition (CVD) diamond detectors have been suggested as ion-beam therapy microdosimeters. Diamond detectors can be manufactured with small cross sections and thin shapes, ideal to cope with the high fluence rate. However the sensitive volume of solid state detectors significantly deviates from conventional microdosimeters, with a diameter that can be up to 1000 times the height. This difference requires a redefinition of the concept of sensitive thickness and a deep study of the secondary to primary radiation, of the wall effects and of the impact of the orientation of the detector with respect to the radiation field. The present work intends to study through Monte Carlo simulations the impact of the detector geometry on the determination of radiation quality quantities, in particular on the relative contribution of primary and secondary radiation. The dependence of microdosimetric quantities such as the unrestricted linear energy L and the lineal energy y are investigated for different detector cross sections, by varying the particle type (carbon ions and protons) and its energy.  
  Address [Solevi, Paola; Magrin, Giulio; Mayer, Ramona] EBG MedAustron, A-2700 Wiener Neustadt, Austria, Email: paola.solevi@ovgu.de  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000361124000004 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 2379  
Permanent link to this record
 

 
Author Piersanti, L.; Bellini, F.; Bini, F.; Collamati, F.; De Lucia, E.; Durante, M.; Faccini, R.; Ferroni, F.; Fiore, S.; Iarocci, E.; La Tessa, C.; Marafini, M.; Mattei, I.; Patera, V.; Ortega, P.G.; Sarti, A.; Schuy, C.; Sciubba, A.; Vanstalle, M.; Voena, C. doi  openurl
  Title Measurement of charged particle yields from PMMA irradiated by a 220 MeV/u C-12 beam Type Journal Article
  Year (down) 2014 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 59 Issue 7 Pages 1857-1872  
  Keywords drift chamber; LYSO; hadrontherapy; carbon ion beam; dose monitoring  
  Abstract The radiation used in hadrontherapy treatments interacts with the patient body producing secondary particles, either neutral or charged, that can be used for dose and Bragg peak monitoring and to provide a fast feedback on the treatment plans. Recent results obtained from the authors on simplified setups (mono-energetic primary beams interacting with homogeneous tissue like target) have already indicated the correlation that exists between the flux of these secondaries coming from the target (e.g. protons and photons) and the position of the primary beam Bragg peak. In this paper, the measurements of charged particle fluxes produced by the interaction of a 220 MeV/u carbon ion beam at GSI, Darmstadt, with a polymethyl methacrylate target are reported. The emission region of protons (p), deuterons (d) and tritons (t) has been characterized using a drift chamber while the particle time-of-flight, used to compute the kinetic energy spectra, was measured with a LYSO scintillator.The energy released in the LYSO crystal was used for particle identification purposes. The measurements were repeated with the setup at 60 degrees and 90 degrees with respect to the primary beam direction. The accuracy on the fragments emission profile reconstruction and its relationship with the Bragg peak position have been studied. Based on the acquired experimental evidence, a method to monitor the dose profile and the position of the Bragg peak inside the target is proposed.  
  Address [Piersanti, L.; De Lucia, E.; Iarocci, E.; Mattei, I.; Sarti, A.] Ist Nazl Fis Nucl, Lab Nazl Frascati, I-00044 Frascati, Italy, Email: vincenzo.patera@lnf.infn.it  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000333186200020 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 1735  
Permanent link to this record
 

 
Author Gillam, J.E.; Solevi, P.; Oliver, J.F.; Casella, C.; Heller, M.; Joram, C.; Rafecas, M. doi  openurl
  Title Sensitivity recovery for the AX-PET prototype using inter-crystal scattering events Type Journal Article
  Year (down) 2014 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 59 Issue 15 Pages 4065-4083  
  Keywords positron emission tomography (PET); inter-crystal scattering; sensitivity  
  Abstract The development of novel detection devices and systems such as the AX-positron emission tomography (PET) demonstrator often introduce or increase the measurement of atypical coincidence events such as inter-crystal scattering (ICS). In more standard systems, ICS events often go undetected and the small measured fraction may be ignored. As the measured quantity of such events in the data increases, so too does the importance of considering them during image reconstruction. Generally, treatment of ICS events will attempt to determine which of the possible candidate lines of response (LoRs) correctly determine the annihilation photon trajectory. However, methods of assessment often have low success rates or are computationally demanding. In this investigation alternative approaches are considered. Experimental data was taken using the AX-PET prototype and a NEMA phantom. Three methods of ICS treatment were assessed-each of which considered all possible candidate LoRs during image reconstruction. Maximum likelihood expectation maximization was used in conjunction with both standard (line-like) and novel (V-like in this investigation) detection responses modeled within the system matrix. The investigation assumed that no information other than interaction locations was available to distinguish between candidates, yet the methods assessed all provided means by which such information could be included. In all cases it was shown that the signal to noise ratio is increased using ICS events. However, only one method, which used full modeling of the ICS response in the system matrix-the V-like model-provided enhancement in all figures of merit assessed in this investigation. Finally, the optimal method of ICS incorporation was demonstrated using data from two small animals measured using the AX-PET demonstrator.  
  Address [Gillam, John E.; Solevi, Paola; Oliver, Josep F.; Rafecas, Magdalena] Univ Valencia, CSIC, IFIC, Valencia, Spain, Email: john.gillam@sydney.edu.au  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000340056800006 Approved no  
  Is ISI yes International Collaboration  
  Call Number IFIC @ pastor @ Serial 1879  
Permanent link to this record
 

 
Author Brzezinski, K.; Oliver, J.F.; Gillam, J.; Rafecas, M. doi  openurl
  Title Study of a high-resolution PET system using a Silicon detector probe Type Journal Article
  Year (down) 2014 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 59 Issue 20 Pages 6117-6140  
  Keywords PET; high-resolution imaging; Si detectors; PET insert  
  Abstract A high-resolution silicon detector probe, in coincidence with a conventional PET scanner, is expected to provide images of higher quality than those achievable using the scanner alone. Spatial resolution should improve due to the finer pixelization of the probe detector, while increased sensitivity in the probe vicinity is expected to decrease noise. A PET-probe prototype is being developed utilizing this principle. The system includes a probe consisting of ten layers of silicon detectors, each a 80 x 52 array of 1 x 1 x 1 mm(3) pixels, to be operated in coincidence with a modern clinical PET scanner. Detailed simulation studies of this system have been performed to assess the effect of the additional probe information on the quality of the reconstructed images. A grid of point sources was simulated to study the contribution of the probe to the system resolution at different locations over the field of view (FOV). A resolution phantom was used to demonstrate the effect on image resolution for two probe positions. A homogeneous source distribution with hot and cold regions was used to demonstrate that the localized improvement in resolution does not come at the expense of the overall quality of the image. Since the improvement is constrained to an area close to the probe, breast imaging is proposed as a potential application for the novel geometry. In this sense, a simplified breast phantom, adjacent to heart and torso compartments, was simulated and the effect of the probe on lesion detectability, through measurements of the local contrast recovery coefficient-to-noise ratio (CNR), was observed. The list-mode ML-EM algorithm was used for image reconstruction in all cases. As expected, the point spread function of the PET-probe system was found to be non-isotropic and vary with position, offering improvement in specific regions. Increase in resolution, of factors of up to 2, was observed in the region close to the probe. Images of the resolution phantom showed visible improvement in resolution when including the probe in the simulations. The image quality study demonstrated that contrast and spill-over ratio in other areas of the FOV were not sacrificed for this enhancement. The CNR study performed on the breast phantom indicates increased lesion detectability provided by the probe.  
  Address [Brzezinski, K.; Oliver, J. F.; Gillam, J.; Rafecas, M.] Univ Valencia, CSIC, Inst Fis Corpuscular, E-46980 Valencia, Spain, Email: brzezinski@ific.uv.es  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000343092300011 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 1963  
Permanent link to this record
 

 
Author Gillam, J.E.; Solevi, P.; Oliver, J.F.; Rafecas, M. doi  openurl
  Title Simulated one-pass list-mode: an approach to on-the-fly system matrix calculation Type Journal Article
  Year (down) 2013 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 58 Issue 7 Pages 2377-2394  
  Keywords  
  Abstract In the development of prototype systems for positron emission tomography a valid and robust image reconstruction algorithm is required. However, prototypes often employ novel detector and system geometries which may change rapidly under optimization. In addition, developing systems generally produce highly granular, or possibly continuous detection domains which require some level of on-the-fly calculation for retention of measurement precision. In this investigation a new method of on-the-fly system matrix calculation is proposed that provides advantages in application to such list-mode systems in terms of flexibility in system modeling. The new method is easily adaptable to complicated system geometries and available computational resources. Detection uncertainty models are used as random number generators to produce ensembles of possible photon trajectories at image reconstruction time for each datum in the measurement list. However, the result of this approach is that the system matrix elements change at each iteration in a non-repetitive manner. The resulting algorithm is considered the simulation of a one-pass list (SOPL) which is generated and the list traversed during image reconstruction. SOPL alters the system matrix in use at each iteration and so behavior within the maximum likelihood-expectation maximization algorithm was investigated. A two-pixel system and a small two dimensional imaging model are used to illustrate the process and quantify aspects of the algorithm. The two-dimensional imaging system showed that, while incurring a penalty in image resolution, in comparison to a non-random equal-computation counterpart, SOPL provides much enhanced noise properties. In addition, enhancement in system matrix quality is straightforward (by increasing the number of samples in the ensemble) so that the resolution penalty can be recovered when desired while retaining improvement in noise properties. Finally the approach is tested and validated against a standard (highly accurate) system matrix using experimental data from a prototype system-the AX-PET.  
  Address [Gillam, J. E.; Solevi, P.; Oliver, J. F.; Rafecas, M.] Univ Valencia, CSIC, IFIC, Inst Fis Corpuscular, Valencia, Spain, Email: john.gillam@ific.uv.es  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000316181300024 Approved no  
  Is ISI yes International Collaboration no  
  Call Number IFIC @ pastor @ Serial 1370  
Permanent link to this record
 

 
Author Robert, C.; Dedes, G.; Battistoni, G.; Bohlen, T.T.; Buvat, I.; Cerutti, F.; Chin, M.P.W.; Ferrari, A.; Gueth, P.; Kurz, C.; Lestand, L.; Mairani, A.; Montarou, G.; Nicolini, R.; Ortega, P.G.; Parodi, K.; Prezado, Y.; Sala, P.R.; Sarrut, D.; Testa, E. doi  openurl
  Title Distributions of secondary particles in proton and carbon-ion therapy: a comparison between GATE/Geant4 and FLUKA Monte Carlo codes Type Journal Article
  Year (down) 2013 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 58 Issue 9 Pages 2879-2899  
  Keywords  
  Abstract Monte Carlo simulations play a crucial role for in-vivo treatment monitoring based on PET and prompt gamma imaging in proton and carbon-ion therapies. The accuracy of the nuclear fragmentation models implemented in these codes might affect the quality of the treatment verification. In this paper, we investigate the nuclear models implemented in GATE/Geant4 and FLUKA by comparing the angular and energy distributions of secondary particles exiting a homogeneous target of PMMA. Comparison results were restricted to fragmentation of O-16 and C-12. Despite the very simple target and set-up, substantial discrepancies were observed between the two codes. For instance, the number of high energy (>1 MeV) prompt gammas exiting the target was about twice as large with GATE/Geant4 than with FLUKA both for proton and carbon ion beams. Such differences were not observed for the predicted annihilation photon production yields, for which ratios of 1.09 and 1.20 were obtained between GATE and FLUKA for the proton beam and the carbon ion beam, respectively. For neutrons and protons, discrepancies from 14% (exiting protons-carbon ion beam) to 57% (exiting neutrons-proton beam) have been identified in production yields as well as in the energy spectra for neutrons.  
  Address Univ Paris 07, IMNC, CNRS, UMR 8165, F-91406 Orsay, France, Email: robert@imnc.in2p3.fr  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000317579900010 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 1407  
Permanent link to this record
 

 
Author Solevi, P. et al; Oliver, J.F.; Gillam, J.E.; Rafecas, M. doi  openurl
  Title A Monte-Carlo based model of the AX-PET demonstrator and its experimental validation Type Journal Article
  Year (down) 2013 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 58 Issue 16 Pages 5495-5510  
  Keywords  
  Abstract AX-PET is a novel PET detector based on axially oriented crystals and orthogonal wavelength shifter (WLS) strips, both individually read out by silicon photo-multipliers. Its design decouples sensitivity and spatial resolution, by reducing the parallax error due to the layered arrangement of the crystals. Additionally the granularity of AX-PET enhances the capability to track photons within the detector yielding a large fraction of inter-crystal scatter events. These events, if properly processed, can be included in the reconstruction stage further increasing the sensitivity. Its unique features require dedicated Monte-Carlo simulations, enabling the development of the device, interpreting data and allowing the development of reconstruction codes. At the same time the non-conventional design of AX-PET poses several challenges to the simulation and modeling tasks, mostly related to the light transport and distribution within the crystals and WLS strips, as well as the electronics readout. In this work we present a hybrid simulation tool based on an analytical model and a Monte-Carlo based description of the AX-PET demonstrator. It was extensively validated against experimental data, providing excellent agreement.  
  Address [Solevi, P.; Oliver, J. F.; Gillam, J. E.; Rafecas, M.] Univ Valencia, CSIC, IFIC, E-46071 Valencia, Spain, Email: paola.solevi@ific.uv.es  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000322775300012 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 1544  
Permanent link to this record
 

 
Author Cabello, J.; Rafecas, M. doi  openurl
  Title Comparison of basis functions for 3D PET reconstruction using a Monte Carlo system matrix Type Journal Article
  Year (down) 2012 Publication Physics in Medicine and Biology Abbreviated Journal Phys. Med. Biol.  
  Volume 57 Issue 7 Pages 1759-1777  
  Keywords  
  Abstract In emission tomography, iterative statistical methods are accepted as the reconstruction algorithms that achieve the best image quality. The accuracy of these methods relies partly on the quality of the system response matrix (SRM) that characterizes the scanner. The more physical phenomena included in the SRM, the higher the SRM quality, and therefore higher image quality is obtained from the reconstruction process. High-resolution small animal scanners contain as many as 10(3)-10(4) small crystal pairs, while the field of view (FOV) is divided into hundreds of thousands of small voxels. These two characteristics have a significant impact on the number of elements to be calculated in the SRM. Monte Carlo (MC) methods have gained popularity as a way of calculating the SRM, due to the increased accuracy achievable, at the cost of introducing some statistical noise and long simulation times. In the work presented here the SRM is calculated using MC methods exploiting the cylindrical symmetries of the scanner, significantly reducing the simulation time necessary to calculate a high statistical quality SRM and the storage space necessary. The use of cylindrical symmetries makes polar voxels a convenient basis function. Alternatively, spherically symmetric basis functions result in improved noise properties compared to cubic and polar basis functions. The quality of reconstructed images using polar voxels, spherically symmetric basis functions on a polar grid, cubic voxels and post-reconstruction filtered polar and cubic voxels is compared from a noise and spatial resolution perspective. This study demonstrates that polar voxels perform as well as cubic voxels, reducing the simulation time necessary to calculate the SRM and the disk space necessary to store it. Results showed that spherically symmetric functions outperform polar and cubic basis functions in terms of noise properties, at the cost of slightly degraded spatial resolution, larger SRM file size and longer reconstruction times. However, we demonstrate that post-reconstruction smoothing, usually applied in emission imaging to reduce the level of noise, can produce a spatial resolution degradation of similar to 50%, while spherically symmetric basis functions produce a degradation of only similar to 6%, compared to polar and cubic voxels, at the same noise level. Therefore, the image quality trade-off obtained with blobs is higher than that obtained with cubic or polar voxels.  
  Address [Cabello, Jorge; Rafecas, Magdalena] Univ Valencia, Inst Fis Corpuscular, CSIC, Valencia, Spain, Email: jorge.cabello@ific.uv.es  
  Corporate Author Thesis  
  Publisher Iop Publishing Ltd Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0031-9155 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000302121000004 Approved no  
  Is ISI yes International Collaboration no  
  Call Number IFIC @ pastor @ Serial 955  
Permanent link to this record