Viegas, R., Roser, J., Barrientos, L., Borja-Lloret, M., Casaña, J. V., Lopez, J. G., et al. (2023). Characterization of a Compton camera based on the TOFPET2 ASIC. Radiat. Phys. Chem., 202, 110507–11pp.
Abstract: The use of Compton cameras for medical imaging and its interest as a hadron therapy treatment monitoring has increased in the last decade with the development of silicon photomultipliers. MACACOp is a Compton camera prototype designed and assembled at the IRIS group of IFIC-Valencia. This Compton camera is based on monolithic Lanthanum (III) Bromide crystals and silicon photomultipliers, and employs the novel TOFPET2 ASIC as readout electronics. This system emerged as an alternative to MACACO II prototype, with the aim of improving its limited time resolution. To test the performance of the ASIC in a Compton camera setup, the prototype was characterized, both in laboratory and in-beam. A time resolution of 1.5 ns was obtained after time corrections, which improves greatly the performance of the MACACO II. Moreover, the results obtained at high photon energies demonstrate the ability of the system to obtain 1 mm displacements of the reconstructed spots. The results reinforce the potential of the system as a monitoring device for hadron therapy.
|
Perez-Calatayud, J., Ballester, F., Tedgren, C., DeWerd, L. A., Papagiannis, P., Rivard, M. J., et al. (2022). GEC-ESTRO ACROP recommendations on calibration and traceability of HE HDR-PDR photon-emitting brachytherapy sources at the hospital level. Radiother. Oncol., 176, 108–117.
Abstract: The vast majority of radiotherapy departments in Europe using brachytherapy (BT) perform temporary implants of high-or pulsed-dose rate (HDR-PDR) sources with photon energies higher than 50 keV. Such techniques are successfully applied to diverse pathologies and clinical scenarios. These recommen-dations are the result of Working Package 21 (WP-21) initiated within the BRAchytherapy PHYsics Quality Assurance System (BRAPHYQS) GEC-ESTRO working group with a focus on HDR-PDR source cal-ibration. They provide guidance on the calibration of such sources, including practical aspects and issues not specifically accounted for in well-accepted societal recommendations, complementing the BRAPHYQS WP-18 Report dedicated to low energy BT photon emitting sources (seeds). The aim of this report is to provide a European-wide standard in HDR-PDR BT source calibration at the hospital level to maintain high quality patient treatments.
|
Jungclaus, A., Doornenbal, P., Acosta, J., Vaquero, V., Browne, F., Cortes, M. L., et al. (2024). Position of the single-particle 3/2- state in 135Sn and the N = 90 subshell closure. Phys. Lett. B, 851, 138561–5pp.
Abstract: The decay of excited states of the nucleus Sn-135, with three neutrons outside the doubly-magic Sn-132 core, was studied in an experiment performed at the Radioactive Isotope Beam Factory at RIKEN. Several gamma rays emitted from excited Sn-135 ions were observed following one-neutron and one-neutron-one-proton removal from Sn-136 and Sb-137 beams, respectively, on a beryllium target at relativistic energies. Based on the analogy to 133Sn populated via one-neutron removal from Sn-134, an excitation energy of 695(15) keV is assigned to the 3/2(-) state with strongest single-particle character in 135Sn. This result provides the first direct information about the evolution of the neutron shell structure beyond N = 82 and thus allows for a crucial test of shellmodel calculations in this region. The experimental findings are in full agreement with calculations performed employing microscopic effective two-body interactions derived from CD-Bonn and N3LO nucleon-nucleon potentials, which do not predict a pronounced subshell gap at neutron number N=90. The occurrence of such a gap in Sn-140, i.e., when the 1f(7/2) orbital is completely filled, had been proposed in the past, in analogy to the magicity of Ca-48, featuring a completely filled 0f(7/2) orbital one harmonic oscillator shell below.
|
Yue, Z. et al, Algora, A., & Nacher, E. (2024). Magnetic moments of thallium isotopes in the vicinity of magic N=126. Phys. Lett. B, 849, 138452–7pp.
Abstract: The magnetic dipole moments (mu) of Tl-209(g)(N =128) and Tl-207(m)(N = 126) have been measured for the first time using the in -source laser resonance -ionization spectroscopy technique with the Laser Ion Source and Trap (LIST) at ISOLDE (CERN). The application of the LIST suppresses the usually overwhelming background of the isobaric francium isotopes and allows access to heavy thallium isotopes with >207. The self -consistent theory of finite Fermi systems based on the energy density functional by Fayans et al. well describes the N dependence of μfor 1/2(+) thallium ground states, as well as μfor the 11/2(-) isomeric states in europium, gold and thallium isotopes. The inclusion of particle-vibration coupling leads to a better agreement between the theory and experiment for mu(Tl-g , I-pi = 1/2(+)). It is shown that beyond mean-field contributions tocannot be neglected at least for thallium isotopes with I-pi = 1/2(+).
|
ATLAS Collaboration(Aad, G. et al), Amos, K. R., Aparisi Pozo, J. A., Bailey, A. J., Bouchhar, N., Cabrera Urban, S., et al. (2024). Observation of Wγγ triboson production in proton-proton collisions at √s=13 TeV with the ATLAS detector. Phys. Lett. B, 848, 138400–24pp.
Abstract: This letter reports the observation of () production in proton-proton collisions. This measurement uses the full Run 2 sample of events recorded at a center-of-mass energy of root= 13 TeV by the ATLAS detector at the LHC, corresponding to an integrated luminosity of 140 fb-1. Events with a leptonically-decaying boson and at least two photons are considered. The background-only hypothesis is rejected with an observed and expected significance of 5.6 standard deviations. The inclusive fiducial production cross section of () and () events is measured to be fid = 13.8 +/- 1.1(stat)+2.1-2.0(syst) +/- 0.1(lumi) fb, in agreement with the Standard Model prediction.
|