Guadilla, V. et al, Tain, J. L., Algora, A., Agramunt, J., Gelletly, W., Jordan, D., et al. (2018). Characterization and performance of the DTAS detector. Nucl. Instrum. Methods Phys. Res. A, 910, 79–89.
Abstract: DTAS is a segmented total absorption y-ray spectrometer developed for the DESPEC experiment at FAIR. It is composed of up to eighteen NaI(Tl) crystals. In this work we study the performance of this detector with laboratory sources and also under real experimental conditions. We present a procedure to reconstruct offline the sum of the energy deposited in all the crystals of the spectrometer, which is complicated by the effect of NaI(Tl) light-yield non-proportionality. The use of a system to correct for time variations of the gain in individual detector modules, based on a light pulse generator, is demonstrated. We describe also an event-based method to evaluate the summing-pileup electronic distortion in segmented spectrometers. All of this allows a careful characterization of the detector with Monte Carlo simulations that is needed to calculate the response function for the analysis of total absorption gamma-ray spectroscopy data. Special attention was paid to the interaction of neutrons with the spectrometer, since they are a source of contamination in studies of beta-delayed neutron emitting nuclei.
|
|
|
Guadilla, V. et al, Algora, A., Tain, J. L., Agramunt, J., Jordan, D., Monserrate, M., et al. (2017). Characterization of a cylindrical plastic beta-detector with Monte Carlo simulations of optical photons. Nucl. Instrum. Methods Phys. Res. A, 854, 134–138.
Abstract: In this work we report on the Monte Carlo study performed to understand and reproduce experimental measurements of a new plastic beta-detector with cylindrical geometry. Since energy deposition simulations differ from the experimental measurements for such a geometry, we show how the simulation of production and transport of optical photons does allow one to obtain the shapes of the experimental spectra. Moreover, taking into account the computational effort associated with this kind of simulation, we develop a method to convert the simulations of energy deposited into light collected, depending only on the interaction point in the detector. This method represents a useful solution when extensive simulations have to be done, as in the case of the calculation of the response function of the spectrometer in a total absorption gamma-ray spectroscopy analysis.
|
|
|
Agramunt, J. et al, Tain, J. L., Albiol, F., Algora, A., Domingo-Pardo, C., Jordan, M. D., et al. (2016). Characterization of a neutron-beta counting system with beta-delayed neutron emitters. Nucl. Instrum. Methods Phys. Res. A, 807, 69–78.
Abstract: A new detection system for the measurement of beta-delayed neutron emission probabilities has been characterized using fission products with well known beta-delayed neutron emission properties. The setup consists of BELEN-20, a 4 pi-neutron counter with twenty He-3 proportional tubes arranged inside a large polyethylene neutron moderator, a thin Si detector for beta counting and a self-triggering digital data acquisition system. The use of delayed-neutron precursors with different neutron emission windows allowed the study of the effect of energy dependency on neutron, beta and beta-neutron rates. The observed effect is well reproduced by Monte Carlo simulations. The impact of this dependency on the accuracy of neutron emission probabilities is discussed. A new accurate value of the neutron emission probability for the important delayed-neutron precursor I-137 was obtained, P-n = 7.76(14)%.
|
|
|
n_TOF Collaboration(Guerrero, C. et al), Giubrone, G., & Tain, J. L. (2011). Characterization of the New n_TOF Neutron Beam: Fluence, Profile and Resolution. J. Korean Phys. Soc., 59(2), 1624–1627.
Abstract: After a halt of four years, the nTOF spallation neutron facility at CERN has resumed operation in November 2008 with a new spallation target characterized by an improved safety and engineering design, resulting in a more robust overall performance and efficient cooling. The first measurement during the 2009 run has aimed at the full characterization of the neutron beam. Several detectors, such as calibrated fission chambers, the nTOF Silicon Monitor, a MicroMegas detector with (10)B and (235)U samples, as well as liquid and solid scintillators have been used in order to characterize the properties of the neutron fluence. The spatial profile of the beam has been studied with a specially designed “X-Y” MicroMegas which provided a 2D image of the beam as a function of neutron energy. Both properties have been compared with simulations performed. with the FLUKA code. The characterization of the resolution function is based on results from simulations which have been verified by the study of narrow capture resonances. of (56)Fe, which were measured as part of a new campaign of (n,gamma) measurements on Fe and Ni isotopes.
|
|
|
BRIKEN Collaboration(Tolosa-Delgado, A. et al), Agramunt, J., Tain, J. L., Algora, A., Domingo-Pardo, C., Morales, A. I., et al. (2019). Commissioning of the BRIKEN detector for the measurement of very exotic beta-delayed neutron emitters. Nucl. Instrum. Methods Phys. Res. A, 925, 133–147.
Abstract: A new detection system has been installed at the RIKEN Nishina Center (Japan) to investigate decay properties of very neutron-rich nuclei. The setup consists of three main parts: a moderated neutron counter, a detection system sensitive to the implantation and decay of radioactive ions, and gamma-ray detectors. We describe here the setup, the commissioning experiment and some selected results demonstrating its performance for the measurement of half-lives and beta-delayed neutron emission probabilities. The methodology followed in the analysis of the data is described in detail. Particular emphasis is placed on the correction of the accidental neutron background.
|
|