AGATA Collaboration(Crespi, F. C. L. et al), & Gadea, A. (2013). Response of AGATA segmented HPGe detectors to gamma rays up to 15.1 MeV. Nucl. Instrum. Methods Phys. Res. A, 705, 47–54.
Abstract: The response of AGATA segmented HPGe detectors to gamma rays in the energy range 2-15 MeV was measured. The 15.1 MeV gamma rays were produced using the reaction d(B-11,n gamma)C-12 at E-beam=19.1 MeV, while gamma rays between 2 and 9 MeV were produced using an Am-Be-Fe radioactive source. The energy resolution and linearity were studied and the energy-to-pulse-height conversion resulted to be linear within 0.05%.Experimental interaction multiplicity distributions are discussed and compared with the results of Geant4 simulations. It is shown that the application of gamma-ray tracking allows a suppression of background radiation caused by n-capture in Ge nuclei. Finally the Doppler correction for the 15.1 MeV gamma line, performed using the position information extracted with Pulse-shape analysis is discussed.
|
|
|
Llosa, G., Barrillon, P., Barrio, J., Bisogni, M. G., Cabello, J., Del Guerra, A., et al. (2013). High performance detector head for PET and PET/MR with continuous crystals and SiPMs. Nucl. Instrum. Methods Phys. Res. A, 702, 3–5.
Abstract: A high resolution PET detector head for small animal PET applications has been developed. The detector is composed of a 12 mm x 12 mm continuous LYSO crystal coupled to a 64-channel monolithic SiPM matrix from FBK-irst. Crystal thicknesses of 5 mm and 10 mm have been tested, both yielding an intrinsic spatial resolution around 0.7 mm FWHM with a position determination algorithm that can also provide depth-of-interaction information. The detectors have been tested in a rotating system that makes it possible to acquire tomographic data and reconstruct images of Na-22 sources. An image reconstruction method specifically adapted for continuous crystals has been employed. The Full Width at Half Maximum measured from a point source reconstructed with ML-EM was 0.7 mm with the 5 mm crystal and 0.8 mm with the 10 mm crystal.
|
|
|
Studen, A., Brzezinski, K., Chesi, E., Cindro, V., Clinthorne, N. H., Cochran, E., et al. (2013). Silicon detectors for combined MR-PET and MR-SPECT imaging. Nucl. Instrum. Methods Phys. Res. A, 702, 88–90.
Abstract: Silicon based devices can extend PET-MR and SPECT-MR imaging to applications, where their advantages in performance outweigh benefits of high statistical counts. Silicon is in many ways an excellent detector material with numerous advantages, among others: excellent energy and spatial resolution, mature processing technology, large signal to noise ratio, relatively low price, availability, versatility and malleability. The signal in silicon is also immune to effects of magnetic field at the level normally used in MR devices. Tests in fields up to 7 T were performed in a study to determine effects of magnetic field on positron range in a silicon PET device. The curvature of positron tracks in direction perpendicular to the field's orientation shortens the distance between emission and annihilation point of the positron. The effect can be fully appreciated for a rotation of the sample for a fixed field direction, compressing range in all dimensions. A popular Ga-68 source was used showing a factor of 2 improvement in image noise compared to zero field operation. There was also a little increase in noise as the reconstructed resolution varied between 2.5 and 1.5 mm. A speculative applications can be recognized in both emission modalities, SPECT and PET. Compton camera is a subspecies of SPECT, where a silicon based scatter as a MR compatible part could inserted into the MR bore and the secondary detector could operate in less constrained environment away from the magnet. Introducing a Compton camera also relaxes requirements of the radiotracers used, extending the range of conceivable photon energies beyond 140.5 keV of the Tc-99m. In PET, one could exploit the compressed sub-millimeter range of positrons in the magnetic field. To exploit the advantage, detectors with spatial resolution commensurate to the effect must be used with silicon being an excellent candidate. Measurements performed outside of the MR achieving spatial resolution below 1 mm are reported.
|
|
|
NEXT Collaboration(Alvarez, V. et al), Carcel, S., Cervera-Villanueva, A., Diaz, J., Ferrario, P., Gil, A., et al. (2013). Near-intrinsic energy resolution for 30-662 keV gamma rays in a high pressure xenon electroluminescent TPC. Nucl. Instrum. Methods Phys. Res. A, 708, 101–114.
Abstract: We present the design, data and results from the NEXT prototype for Double Beta and Dark Matter (NEXT-DBDM) detector, a high-pressure gaseous natural xenon electroluminescent time projection chamber (TPC) that was built at the Lawrence Berkeley National Laboratory. It is a prototype of the planned NEXT-100 Xe-136 neutrino-less double beta decay (0 nu beta beta) experiment with the main objectives of demonstrating near-intrinsic energy resolution at energies up to 662 keV and of optimizing the NEXT-100 detector design and operating parameters. Energy resolutions of similar to 1% FWHM for 662 keV gamma rays were obtained at 10 and 15 atm and similar to 5% FWHM for 30 keV fluorescence xenon X-rays. These results demonstrate that 0.5% FWHM resolutions for the 2459 keV hypothetical neutrino-less double beta decay peak are realizable. This energy resolution is a factor 7-20 better than that of the current leading 0 nu beta beta experiments using liquid xenon and thus represents a significant advancement. We present also first results from a track imaging system consisting of 64 silicon photo-multipliers recently installed in NEXT-DBDM that, along with the excellent energy resolution, demonstrates the key functionalities required for the NEXT-100 0 nu beta beta search.
|
|
|
BABAR Collaboration(Lees, J. P. et al), Martinez-Vidal, F., Oyanguren, A., & Villanueva-Perez, P. (2013). Time-integrated luminosity recorded by the BABAR detector at the PEP-II e(+)e(-) collider. Nucl. Instrum. Methods Phys. Res. A, 726, 203–213.
Abstract: We describe a measurement of the time-integrated luminosity of the data collected by the BABAR experiment at the PEP-II asymmetric-energy e(+)e(-) collider at the Upsilon(4S), Upsilon(3S), and Upsilon(2S) resonances and in a continuum region below each resonance. We measure the time-integrated luminosity by counting e(+)e(-)-> e(+)e(-) and (for the Upsilon(4S) only) e(+)e(-)->mu(+)mu(-) candidate events, allowing additional photons in the final state. We use data-corrected simulation to determine the cross-sections and reconstruction efficiencies for these processes, as well as the major backgrounds. Due to the large cross-sections of e(+)e(-)-> e(+)e(-) and e(+)e(-)->mu(+)mu(-), the statistical uncertainties of the measurement are substantially smaller than the systematic uncertainties. The dominant systematic uncertainties are due to observed differences between data and simulation, as well as uncertainties on the cross-sections. For data collected on the Upsilon(3S) and Upsilon(2S) resonances, an additional uncertainty arises due to Upsilon -> e(+)e(-)X background. For data collected off the Upsilon resonances, we estimate an additional uncertainty due to time dependent efficiency variations, which can affect the short off-resonance runs. The relative uncertainties on the luminosities of the on-resonance (off-resonance) samples are 0.43% (0.43%) for the Upsilon(4S), 0.58% (0.72%) for the Upsilon(3S), and 0.68% (0.88%) for the Upsilon(2S).
|
|