Records |
Links |
|
Author |
Beltrame, P. et al; Oliver, J.F.; Rafecas, M.; Solevi, P. |
![find record details (via OpenURL) openurl](img/xref.gif)
|
|
Title |
The AX-PET demonstrator-Design, construction and characterization |
Type |
Journal Article |
|
Year |
2011 |
Publication |
Nuclear Instruments & Methods in Physics Research A |
Abbreviated Journal |
Nucl. Instrum. Methods Phys. Res. A |
|
|
Volume |
654 |
Issue ![sorted by Issue field, ascending order (up)](img/sort_asc.gif) |
1 |
Pages |
546-559 |
|
|
Keywords |
PET; Axial geometry; Geiger-mode Avalanche Photo Diodes (G-APD); SiPM |
|
|
Abstract |
Axial PET is a novel geometrical concept for Positron Emission Tomography (PET), based on layers of long scintillating crystals axially aligned with the bore axis. The axial coordinate is obtained from arrays of wavelength shifting (WLS) plastic strips placed orthogonally to the crystals. This article describes the design, construction and performance evaluation of a demonstrator set-up which consists of two identical detector modules, used in coincidence. Each module comprises 48 LYSO crystals of 100 mm length and 156 WLS strips. Crystals and strips are readout by Geiger-mode Avalanche Photo Diodes (G-APDs). The signals from the two modules are processed by fully analog front-end electronics and recorded in coincidence by a VME-based data acquisition system. Measurements with point-like (22)Na sources, with the modules used both individually and in coincidence mode, allowed for a complete performance evaluation up to the focal plane reconstruction of point sources. The results obtained are in good agreement with expectations and proved the set-up to be ready for the next evaluation phase with PET phantoms filled with radiotracers. |
|
|
Address |
[Casella, C; Dissertori, G; Djambazov, L; Lustermann, W; Nessi-Tedaldi, F; Pauss, F; Renker, D; Schinzel, D] ETH, Inst Particle Phys, CH-8093 Zurich, Switzerland, Email: Chiara.Casella@cern.ch |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Elsevier Science Bv |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0168-9002 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
WOS:000295765100078 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
yes |
|
|
Call Number |
IFIC @ pastor @ |
Serial |
790 |
|
Permanent link to this record |
|
|
|
|
Author |
Aguiar, P.; Rafecas, M.; Ortuño, J.E.; Kontaxakis, G.; Santos, A.; Pavia, J.; Rosetti, M. |
![find record details (via OpenURL) openurl](img/xref.gif)
|
|
Title |
Geometrical and Monte Carlo projectors in 3D PET reconstruction |
Type |
Journal Article |
|
Year |
2010 |
Publication |
Medical Physics |
Abbreviated Journal |
Med. Phys. |
|
|
Volume |
37 |
Issue ![sorted by Issue field, ascending order (up)](img/sort_asc.gif) |
11 |
Pages |
5691-5702 |
|
|
Keywords |
3D PET; iterative reconstruction; list-mode reconstruction; ray-tracing techniques; Monte Carlo simulation; system response matrix |
|
|
Abstract |
Purpose: In the present work, the authors compare geometrical and Monte Carlo projectors in detail. The geometrical projectors considered were the conventional geometrical Siddon ray-tracer (S-RT) and the orthogonal distance-based ray-tracer (OD-RT), based on computing the orthogonal distance from the center of image voxel to the line-of-response. A comparison of these geometrical projectors was performed using different point spread function (PSF) models. The Monte Carlo-based method under consideration involves an extensive model of the system response matrix based on Monte Carlo simulations and is computed off-line and stored on disk. Methods: Comparisons were performed using simulated and experimental data of the commercial small animal PET scanner rPET. Results: The results demonstrate that the orthogonal distance-based ray-tracer and Siddon ray-tracer using PSF image-space convolutions yield better images in terms of contrast and spatial resolution than those obtained after using the conventional method and the multiray-based S-RT. Furthermore, the Monte Carlo-based method yields slight improvements in terms of contrast and spatial resolution with respect to these geometrical projectors. Conclusions: The orthogonal distance-based ray-tracer and Siddon ray-tracer using PSF image-space convolutions represent satisfactory alternatives to factorizing the system matrix or to the conventional on-the-fly ray-tracing methods for list-mode reconstruction, where an extensive modeling based on Monte Carlo simulations is unfeasible. |
|
|
Address |
[Aguiar, Pablo] Univ Santiago de Compostela, Dept Fis Particulas, Complexo Hosp Univ Santiago de Compostela, Fdn IDICHUS IDIS, Santiago De Compostela, Spain, Email: pablo.aguiar.fernandez@sergas.es |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Amer Assoc Physicists Medicine Amer Inst Physics |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0094-2405 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
ISI:000283747600015 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
no |
|
|
Call Number |
IFIC @ elepoucu @ |
Serial |
338 |
|
Permanent link to this record |
|
|
|
|
Author |
Blume, M.; Martinez-Moller, A.; Keil, A.; Navab, N.; Rafecas, M. |
![find record details (via OpenURL) openurl](img/xref.gif)
|
|
Title |
Joint Reconstruction of Image and Motion in Gated Positron Emission Tomography |
Type |
Journal Article |
|
Year |
2010 |
Publication |
IEEE Transactions on Medical Imaging |
Abbreviated Journal |
IEEE Trans. Med. Imaging |
|
|
Volume |
29 |
Issue ![sorted by Issue field, ascending order (up)](img/sort_asc.gif) |
11 |
Pages |
1892-1906 |
|
|
Keywords |
Gating; motion compensation; positron emission tomography (PET); reconstruction |
|
|
Abstract |
We present a novel intrinsic method for joint reconstruction of both image and motion in positron emission tomography (PET). Intrinsic motion compensation methods exclusively work on the measured data, without any external motion measurements. Most of these methods separate image from motion estimation: They use deformable image registration/optical flow techniques in order to estimate the motion from individually reconstructed gates. Then, the image is estimated based on this motion information. With these methods, a main problem lies in the motion estimation step, which is based on the noisy gated frames. The more noise is present, the more inaccurate the image registration becomes. As we show both visually and quantitatively, joint reconstruction using a simple deformation field motion model can compete with state-of-the-art image registration methods which use robust multilevel B-spline motion models. |
|
|
Address |
[Blume, Moritz; Rafecas, Magdalena] Univ Valencia, CSIC, IFIC, E-46071 Valencia, Spain, Email: moritz.blume@cs.tum.edu |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Ieee-Inst Electrical Electronics Engineers Inc |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0278-0062 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
ISI:000283941800007 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
yes |
|
|
Call Number |
IFIC @ elepoucu @ |
Serial |
340 |
|
Permanent link to this record |
|
|
|
|
Author |
Solevi, P.; Muñoz, E.; Solaz, C.; Trovato, M.; Dendooven, P.; Gillam, J.E.; Lacasta, C.; Oliver, J.F.; Rafecas, M.; Torres-Espallardo, I.; Llosa, G. |
![find record details (via OpenURL) openurl](img/xref.gif)
|
|
Title |
Performance of MACACO Compton telescope for ion-beam therapy monitoring: first test with proton beams |
Type |
Journal Article |
|
Year |
2016 |
Publication |
Physics in Medicine and Biology |
Abbreviated Journal |
Phys. Med. Biol. |
|
|
Volume |
61 |
Issue ![sorted by Issue field, ascending order (up)](img/sort_asc.gif) |
14 |
Pages |
5149-5165 |
|
|
Keywords |
ion-beam therapy; range verification; prompt gamma; Compton camera; GATE |
|
|
Abstract |
In order to exploit the advantages of ion-beam therapy in a clinical setting, delivery verification techniques are necessary to detect deviations from the planned treatment. Efforts are currently oriented towards the development of devices for real-time range monitoring. Among the different detector concepts proposed, Compton cameras are employed to detect prompt gammas and represent a valid candidate for real-time range verification. We present the first on-beam test of MACACO, a Compton telescope (multi-layer Compton camera) based on lanthanum bromide crystals and silicon photo-multipliers. The Compton telescope was first characterized through measurements and Monte Carlo simulations. The detector linearity was measured employing Na-22 and Am-Be sources, obtaining about 10% deviation from linearity at 3.44 MeV. A spectral image reconstruction algorithm was tested on synthetic data. Point-like sources emitting gamma rays with energy between 2 and 7 MeV were reconstructed with 3-5 mm resolution. The two-layer Compton telescope was employed to measure radiation emitted from a beam of 150 MeV protons impinging on a cylindrical PMMA target. Bragg-peak shifts were achieved via adjustment of the PMMA target location and the resulting measurements used during image reconstruction. Reconstructed Bragg peak profiles proved sufficient to observe peak-location differences within 10 mm demonstrating the potential of the MACACO Compton Telescope as a monitoring device for ion-beam therapy. |
|
|
Address |
[Solevi, Paola; Munoz, Enrique; Solaz, Carles; Trovato, Marco; Gillam, John E.; Lacasta, Carlos; Oliver, Josep F.; Rafecas, Magdalena; Torres-Espallardo, Irene; Llosa, Gabriela] IFIC CSIC UVEG, Inst Fis Corpuscular, Valencia, Spain, Email: paola.solevi@ovgu.de |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Iop Publishing Ltd |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0031-9155 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
WOS:000379555300007 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
yes |
|
|
Call Number |
IFIC @ pastor @ |
Serial |
2754 |
|
Permanent link to this record |
|
|
|
|
Author |
Gillam, J.E.; Solevi, P.; Oliver, J.F.; Casella, C.; Heller, M.; Joram, C.; Rafecas, M. |
![find record details (via OpenURL) openurl](img/xref.gif)
|
|
Title |
Sensitivity recovery for the AX-PET prototype using inter-crystal scattering events |
Type |
Journal Article |
|
Year |
2014 |
Publication |
Physics in Medicine and Biology |
Abbreviated Journal |
Phys. Med. Biol. |
|
|
Volume |
59 |
Issue ![sorted by Issue field, ascending order (up)](img/sort_asc.gif) |
15 |
Pages |
4065-4083 |
|
|
Keywords |
positron emission tomography (PET); inter-crystal scattering; sensitivity |
|
|
Abstract |
The development of novel detection devices and systems such as the AX-positron emission tomography (PET) demonstrator often introduce or increase the measurement of atypical coincidence events such as inter-crystal scattering (ICS). In more standard systems, ICS events often go undetected and the small measured fraction may be ignored. As the measured quantity of such events in the data increases, so too does the importance of considering them during image reconstruction. Generally, treatment of ICS events will attempt to determine which of the possible candidate lines of response (LoRs) correctly determine the annihilation photon trajectory. However, methods of assessment often have low success rates or are computationally demanding. In this investigation alternative approaches are considered. Experimental data was taken using the AX-PET prototype and a NEMA phantom. Three methods of ICS treatment were assessed-each of which considered all possible candidate LoRs during image reconstruction. Maximum likelihood expectation maximization was used in conjunction with both standard (line-like) and novel (V-like in this investigation) detection responses modeled within the system matrix. The investigation assumed that no information other than interaction locations was available to distinguish between candidates, yet the methods assessed all provided means by which such information could be included. In all cases it was shown that the signal to noise ratio is increased using ICS events. However, only one method, which used full modeling of the ICS response in the system matrix-the V-like model-provided enhancement in all figures of merit assessed in this investigation. Finally, the optimal method of ICS incorporation was demonstrated using data from two small animals measured using the AX-PET demonstrator. |
|
|
Address |
[Gillam, John E.; Solevi, Paola; Oliver, Josep F.; Rafecas, Magdalena] Univ Valencia, CSIC, IFIC, Valencia, Spain, Email: john.gillam@sydney.edu.au |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Iop Publishing Ltd |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0031-9155 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
WOS:000340056800006 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
|
|
|
Call Number |
IFIC @ pastor @ |
Serial |
1879 |
|
Permanent link to this record |