Babiano, V., Balibrea, J., Caballero, L., Calvo, D., Ladarescu, I., Mira Prats, S., et al. (2020). First i-TED demonstrator: A Compton imager with Dynamic Electronic Collimation. Nucl. Instrum. Methods Phys. Res. A, 953, 163228–9pp.
Abstract: i-TED consists of both a total energy detector and a Compton camera primarily intended for the measurement of neutron capture cross sections by means of the simultaneous combination of neutron time-of-flight (TOF) and gamma-ray imaging techniques. TOF allows one to obtain a neutron-energy differential capture yield, whereas the imaging capability is intended for the discrimination of radiative background sources, that have a spatial origin different from that of the capture sample under investigation. A distinctive feature of i-TED is the embedded Dynamic Electronic Collimation (DEC) concept, which allows for a trade-off between efficiency and image resolution. Here we report on some general design considerations and first performance characterization measurements made with an i-TED demonstrator in order to explore its gamma-ray detection and imaging capabilities.
|