Author (up) Studen, A.; Burdette, D.; Chesi, E.; Cindro, V.; Clinthorne, N.H.; Cochran, E.; Grosicar, B.; Kagan, H.; Lacasta, C.; Linhart, V.; Mikuz, M.; Stankova, V.; Weilhammer, P.; Zontar, D. doi  openurl
  Title Timing performance of the silicon PET insert probe Type Journal Article
  Year 2010 Publication Radiation Protection Dosimetry Abbreviated Journal Radiat. Prot. Dosim.  
  Volume 139 Issue 1-3 Pages 199-203  
  Keywords  
  Abstract Simulation indicates that PET image could be improved by upgrading a conventional ring with a probe placed close to the imaged object. In this paper, timing issues related to a PET probe using high-resistivity silicon as a detector material are addressed. The final probe will consist of several (four to eight) 1-mm thick layers of silicon detectors, segmented into 1 x 1 mm(2) pads, each pad equivalent to an independent p + nn+ diode. A proper matching of events in silicon with events of the external ring can be achieved with a good timing resolution. To estimate the timing performance, measurements were performed on a simplified model probe, consisting of a single 1-mm thick detector with 256 square pads (1.4 mm side), coupled with two VATAGP7s, application-specific integrated circuits. The detector material and electronics are the same that will be used for the final probe. The model was exposed to 511 keV annihilation photons from an Na-22 source, and a scintillator (LYSO)-PMT assembly was used as a timing reference. Results were compared with the simulation, consisting of four parts: (i) GEANT4 implemented realistic tracking of electrons excited by annihilation photon interactions in silicon, (ii) calculation of propagation of secondary ionisation (electron-hole pairs) in the sensor, (iii) estimation of the shape of the current pulse induced on surface electrodes and (iv) simulation of the first electronics stage. A very good agreement between the simulation and the measurements were found. Both indicate reliable performance of the final probe at timing windows down to 20 ns.  
  Address [Studen, A.; Cindro, V.; Grosicar, B.; Mikuz, M.; Zontar, D.] Jozef Stefan Inst, SI-1000 Ljubljana, Slovenia, Email: andrej.studen@ijs.si  
  Corporate Author Thesis  
  Publisher Oxford Univ Press Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0144-8420 ISBN Medium  
  Area Expedition Conference  
  Notes ISI:000277738200035 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ elepoucu @ Serial 449  
Permanent link to this record
 

 
Author (up) Studen, A.; Chesi, E.; Cindro, V.; Clinthorne, N.H.; Cochran, E.; Grosicar, B.; Honscheid, K.; Kagan, H.; Lacasta, C.; Llosa, G.; Linhart, V.; Mikuz, M.; Stankova, V.; Weilhammer, P.; Zontar, D. doi  openurl
  Title A silicon PET probe Type Journal Article
  Year 2011 Publication Nuclear Instruments & Methods in Physics Research A Abbreviated Journal Nucl. Instrum. Methods Phys. Res. A  
  Volume 648 Issue Pages S255-S258  
  Keywords PET; Silicon detectors  
  Abstract PET scanners with high spatial resolution offer a great potential in improving diagnosis, therapy monitoring and treatment validation for several severe diseases. One way to improve resolution of a PET scanner is to extend a conventional PET ring with a small probe with excellent spatial resolution. The probe is intended to be placed close to the area of interest. The coincidences of interactions within the probe and the external ring provide a subset of data which combined with data from external ring, greatly improve resolution in the area viewed by the probe. Our collaboration is developing a prototype of a PET probe, composed of high-resolution silicon pad detectors. The detectors are 1 mm thick, measuring 40 by 26 mm(2), and several such sensors are envisaged to either compensate for low stopping power of silicon or increase the area covered by the probe. The sensors are segmented into 1 mm(3) cubic voxels, giving 1040 readout pads per sensor. A module is composed of two sensors placed in a back-to-back configuration, allowing for stacking fraction of up to 70% within a module. The pads are coupled to a set of 16 ASICs (VaTaGP7.1 by IDEAS) per module and read out through a custom designed data acquisition board, allowing for trigger and data interfacing with the external ring. This paper presents an overview of probe requirements and expected performance parameters. It will focus on the characteristics of the silicon modules and their impact on overall probe performance, including spatial resolution, energy resolution and timing resolution. We will show that 1 mm(3) voxels will significantly extend the spatial resolution of conventional PET rings, and that broadening of timing resolution related to varying depth of photon interactions can be compensated to match the timing resolution of the external ring. The initial test results of the probe will also be presented.  
  Address [Studen, A.; Cindro, V.; Grosicar, B.; Mikuz, M.; Zontar, D.] Jozef Stefan Inst, Ljubljana, Slovenia, Email: andrej.studen@ijs.si  
  Corporate Author Thesis  
  Publisher Elsevier Science Bv Place of Publication Editor  
  Language English Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN 0168-9002 ISBN Medium  
  Area Expedition Conference  
  Notes WOS:000305376900063 Approved no  
  Is ISI yes International Collaboration yes  
  Call Number IFIC @ pastor @ Serial 1070  