Gomez-Cadenas, J. J., Benlloch-Rodriguez, J. M., & Ferrario, P. (2016). Application of scintillating properties of liquid xenon and silicon photomultiplier technology to medical imaging. Spectroc. Acta Pt. B, 118, 6–13.
Abstract: We describe a new positron emission time-of-flight apparatus using liquid xenon. The detector is based in a liquid xenon scintillating cell. The cell shape and dimensions can be optimized depending on the intended application. In its simplest form, the liquid xenon scintillating cell is a box in which two faces are covered by silicon photomultipliers and the others by a reflecting material such as Teflon. It is a compact, homogenous and highly efficient detector which shares many of the desirable properties of monolithic crystals, with the added advantage of high yield and fast scintillation offered by liquid xenon. Our initial studies suggest that good energy and spatial resolution comparable with that achieved by lutetium oxyorthosilicate crystals can be obtained with a detector based in liquid xenon scintillating cells. In addition, the system can potentially achieve an excellent coincidence resolving time of better than 100 ps.
|
|
|
Gomez-Cadenas, J. J., Benlloch-Rodriguez, J. M., & Ferrario, P. (2017). Monte Carlo study of the coincidence resolving time of a liquid xenon PET scanner, using Cherenkov radiation. J. Instrum., 12, P08023–13pp.
Abstract: In this paper we use detailed Monte Carlo simulations to demonstrate that liquid xenon (LXe) can be used to build a Cherenkov-based TOF-PET, with an intrinsic coincidence resolving time (CRT) in the vicinity of 10 ps. This extraordinary performance is due to three facts: a) the abundant emission of Cherenkov photons by liquid xenon; b) the fact that LXe is transparent to Cherenkov light; and c) the fact that the fastest photons in LXe have wavelengths higher than 300 nm, therefore making it possible to separate the detection of scintillation and Cherenkov light. The CRT in a Cherenkov LXe TOF-PET detector is, therefore, dominated by the resolution (time jitter) introduced by the photosensors and the electronics. However, we show that for sufficiently fast photosensors (e.g, an overall 40 ps jitter, which can be achieved by current micro-channel plate photomultipliers) the overall CRT varies between 30 and 55 ps, depending on the detection efficiency. This is still one order of magnitude better than commercial CRT devices and improves by a factor 3 the best CRT obtained with small laboratory prototypes.
|
|
|
Gomez-Cadenas, J. J., Benlloch-Rodriguez, J. M., Ferrario, P., Monrabal, F., Rodriguez, J., & Toledo, J. F. (2016). Investigation of the coincidence resolving time performance of a PET scanner based on liquid xenon: a Monte Carlo study. J. Instrum., 11, P09011–18pp.
Abstract: The measurement of the time of flight of the two 511 keV gammas recorded in coincidence in a PET scanner provides an effective way of reducing the random background and therefore increases the scanner sensitivity, provided that the coincidence resolving time (CRT) of the gammas is sufficiently good. The best commercial PET-TOF system today (based in LYSO crystals and digital SiPMs), is the VEREOS of Philips, boasting a CRT of 316 ps (FWHM). In this paper we present a Monte Carlo investigation of the CRT performance of a PET scanner exploiting the scintillating properties of liquid xenon. We find that an excellent CRT of 70 ps (depending on the PDE of the sensor) can be obtained if the scanner is instrumented with silicon photomultipliers (SiPMs) sensitive to the ultraviolet light emitted by xenon. Alternatively, a CRT of 160 ps can be obtained instrumenting the scanner with (much cheaper) blue-sensitive SiPMs coated with a suitable wavelength shifter. These results show the excellent time of flight capabilities of a PET device based in liquid xenon.
|
|
|
Grkovski, M., Brzezinski, K., Cindro, V., Clinthorne, N. H., Kagan, H., Lacasta, C., et al. (2015). Evaluation of a high resolution silicon PET insert module. Nucl. Instrum. Methods Phys. Res. A, 788, 86–94.
Abstract: Conventional PET systems can be augmented with additional detectors placed in close proximity of the region of interest. We developed a high resolution PET insert module to evaluate the added benefit of such a combination. The insert module consists of two back-to-back 1 mm thick silicon sensors, each segmented into 1040 1 mm(2) pads arranged in a 40 by 26 array. A set of 16 VATAGP7.1 ASICs and a custom assembled data acquisition board were used to read out the signal from the insert module. Data were acquired in slice (20) geometry with a Jaszczak phantom (rod diameters of 12-4.8 mm) Filled with F-18-FDG and the images were reconstructed with ML-EM method. Both data with full and limited angular coverage from the insert module were considered and three types of coincidence events were combined. The ratio of high-resolution data that substantially improves quality of the reconstructed image for the region near the surface of the insert module was estimated to be about 4%. Results from our previous studies suggest that such ratio could be achieved at a moderate technological expense by using an equivalent of two insert modules (an effective sensor thickness of 4 mm).
|
|
|
Koolen, B. B., Vidal-Sicart, S., Benlloch, J. M., & Olmos, R. A. V. (2014). Evaluating heterogeneity of primary tumor 18 F-FDG uptake in breast cancer with a dedicated breast PET ( MAMMI): a feasibility study based on correlation with PET/CT. Nucl. Med. Commun., 35(5), 446–452.
Abstract: PurposeThe aim of the study was to evaluate the heterogeneity of primary tumor F-18-fluorodeoxyglucose (F-18-FDG) uptake in breast cancer patients using a dedicated breast PET.Patients and methodsA positron emission tomography/computed tomography (PET/CT) of the thorax was performed 60 min after administration of 180-240 MBq of F-18-FDG in patients with breast cancer. Subsequently, 110 min after injection, a scan was taken with a dedicated high-resolution breast PET [MAMmography with Molecular Imaging (MAMMI)]. Both procedures were performed with the patients in the prone position. Four-point scores were used to compare the intensity (0: none; 1: mild; 2: moderate; 3: high) and heterogeneity (0: none; 1: mild; 2: moderate; 3: high) of F-18-FDG uptake between PET/CT and MAMMI images.ResultsThirty-five patients in whom the primary tumor was visualized on both scans were included in this analysis. The mean primary tumor size was 35.1 mm (range 10-108 mm). The mean intensity score was similar on both devices (2.4 for PET/CT and 2.3 for MAMMI; P=0.439), but the mean heterogeneity score on MAMMI images was significantly higher (PET/CT 1.9 vs. MAMMI 2.3; P=0.005). MAMMI showed a higher heterogeneity score in 11 (31%) of 35 patients, especially in tumors with moderate or high intensity. Significantly higher heterogeneity scores on both PET/CT and MAMMI were seen in large tumors (P=0.005 and 0.014, respectively) and in tumors with high intensity scores (P=0.012 and P<0.001, respectively).ConclusionHeterogeneous tumor F-18-FDG uptake in breast cancer is frequently observed, particularly in large tumors with intense F-18-FDG uptake. It is more often seen on MAMMI PET than on conventional PET/CT. Although the observed heterogeneity should be proven histopathologically, this finding offers a rationale for F-18-FDG-guided biopsies.
|
|