Beltrame, P. et al, Oliver, J. F., Rafecas, M., & Solevi, P. (2011). The AX-PET demonstrator-Design, construction and characterization. Nucl. Instrum. Methods Phys. Res. A, 654(1), 546–559.
Abstract: Axial PET is a novel geometrical concept for Positron Emission Tomography (PET), based on layers of long scintillating crystals axially aligned with the bore axis. The axial coordinate is obtained from arrays of wavelength shifting (WLS) plastic strips placed orthogonally to the crystals. This article describes the design, construction and performance evaluation of a demonstrator set-up which consists of two identical detector modules, used in coincidence. Each module comprises 48 LYSO crystals of 100 mm length and 156 WLS strips. Crystals and strips are readout by Geiger-mode Avalanche Photo Diodes (G-APDs). The signals from the two modules are processed by fully analog front-end electronics and recorded in coincidence by a VME-based data acquisition system. Measurements with point-like (22)Na sources, with the modules used both individually and in coincidence mode, allowed for a complete performance evaluation up to the focal plane reconstruction of point sources. The results obtained are in good agreement with expectations and proved the set-up to be ready for the next evaluation phase with PET phantoms filled with radiotracers.
|
Cabanelas, P. et al, & Nacher, E. (2020). Performance recovery of long CsI(Tl) scintillator crystals with APD-based readout. Nucl. Instrum. Methods Phys. Res. A, 965, 163845–6pp.
Abstract: CALIFA is the high efficiency and energy resolution calorimeter for the (RB)-B-3 experiment at FAIR, intended for detecting high energy light charged particles and gamma rays in scattering experiments, and is being commissioned during the Phase-0 experiments at FAIR, between 2018 and 2020. It surrounds the reaction target in a segmented configuration with 2432 detection units made of long CsI(Tl) finger-shaped scintillator crystals. CALIFA has a 10 year intended operational lifetime as the (RB)-B-3 calorimeter, necessitating measures to be taken to ensure enduring performance. In this paper we present a systematic study of two groups of 6 different detection units of the CALIFA detector after more than four years of operation. The energy resolution and light output yield are evaluated under different conditions. Tests cover the aging of the first detector units assembled and investigates recovery procedures for degraded detection units. A possible reason for the observed degradation is given, pointing to the crystal-APD coupling.
|
n_TOF Collaboration(Balibrea-Correa, J. et al), Lerendegui-Marco, J., Domingo-Pardo, C., Ladarescu, I., Tarifeño-Saldivia, A., de la Fuente-Rosales, G., et al. (2025). Towards a new generation of solid total-energy detectors for neutron-capture time-of-flight experiments with intense neutron beams. Nucl. Instrum. Methods Phys. Res. A, 1072, 170110–14pp.
Abstract: Challenging neutron-capture cross-section measurements of small cross sections and samples with a very limited number of atoms require high-flux time-of-flight facilities. In turn, such facilities need innovative detection setups that are fast, have low sensitivity to neutrons, can quickly recover from the so-called. gamma-flash, and offer the highest possible detection sensitivity. In this paper, we present several steps towards such advanced systems. Specifically, we describe the performance of a high-sensitivity experimental setupat CERN n_TOF EAR2. It consists of nine sTED detector modules in a compact cylindrical configuration, two conventional used large-volume C6D6 detectors, and one LaCl3(Ce) detector. The performance of these detection systems is compared using Nb-93(n, gamma) data. We also developed a detailed GEANT4 Monte Carlo model of the experimental EAR2 setup, which allows for a better understanding of the detector features, including their efficiency determination. This Monte Carlo model has been used for further optimization, thus leading to a new conceptual design of a gamma detector array, STAR, based on a deuterated-stilbene crystal array. Finally, the suitability of deuterated-stilbene crystals for the future STAR array is investigated experimentally utilizing a small stilbene-d12 prototype. The results suggest a similar or superior performance of STAR with respect to other setups based on liquid-scintillators, and allow for additional features such as neutron-gamma discrimination and a higher level of customization capability.
|
Llosa, G. (2019). SiPM-based Compton cameras. Nucl. Instrum. Methods Phys. Res. A, 926, 148–152.
Abstract: Compton cameras have been developed for almost fifty years in various fields (astronomy, medical imaging, safety and industrial inspections, etc.), employing different types of detectors. Their potential use has gained renewed interest with the emergence of high light yield scintillator crystals and silicon photomultipliers (SiPMs). This combination provides good performance and operation simplicity at an affordable cost, raising again the interest in this type of systems. SiPM-based Compton cameras are being assessed for diverse applications with promising results.
|
Grkovski, M., Brzezinski, K., Cindro, V., Clinthorne, N. H., Kagan, H., Lacasta, C., et al. (2015). Evaluation of a high resolution silicon PET insert module. Nucl. Instrum. Methods Phys. Res. A, 788, 86–94.
Abstract: Conventional PET systems can be augmented with additional detectors placed in close proximity of the region of interest. We developed a high resolution PET insert module to evaluate the added benefit of such a combination. The insert module consists of two back-to-back 1 mm thick silicon sensors, each segmented into 1040 1 mm(2) pads arranged in a 40 by 26 array. A set of 16 VATAGP7.1 ASICs and a custom assembled data acquisition board were used to read out the signal from the insert module. Data were acquired in slice (20) geometry with a Jaszczak phantom (rod diameters of 12-4.8 mm) Filled with F-18-FDG and the images were reconstructed with ML-EM method. Both data with full and limited angular coverage from the insert module were considered and three types of coincidence events were combined. The ratio of high-resolution data that substantially improves quality of the reconstructed image for the region near the surface of the insert module was estimated to be about 4%. Results from our previous studies suggest that such ratio could be achieved at a moderate technological expense by using an equivalent of two insert modules (an effective sensor thickness of 4 mm).
|