Records |
Links |
|
Author |
Grkovski, M.; Brzezinski, K.; Cindro, V.; Clinthorne, N.H.; Kagan, H.; Lacasta, C.; Mikuz, M.; Solaz, C.; Studen, A.; Weilhammer, P.; Zontar, D. |
|
|
Title |
Evaluation of a high resolution silicon PET insert module |
Type |
Journal Article |
|
Year |
2015 |
Publication |
Nuclear Instruments & Methods in Physics Research A |
Abbreviated Journal |
Nucl. Instrum. Methods Phys. Res. A |
|
|
Volume |
788 |
Issue |
|
Pages |
86-94 |
|
|
Keywords |
Positron emission tomography; Silicon detectors; PET insert; Image reconstruction |
|
|
Abstract |
Conventional PET systems can be augmented with additional detectors placed in close proximity of the region of interest. We developed a high resolution PET insert module to evaluate the added benefit of such a combination. The insert module consists of two back-to-back 1 mm thick silicon sensors, each segmented into 1040 1 mm(2) pads arranged in a 40 by 26 array. A set of 16 VATAGP7.1 ASICs and a custom assembled data acquisition board were used to read out the signal from the insert module. Data were acquired in slice (20) geometry with a Jaszczak phantom (rod diameters of 12-4.8 mm) Filled with F-18-FDG and the images were reconstructed with ML-EM method. Both data with full and limited angular coverage from the insert module were considered and three types of coincidence events were combined. The ratio of high-resolution data that substantially improves quality of the reconstructed image for the region near the surface of the insert module was estimated to be about 4%. Results from our previous studies suggest that such ratio could be achieved at a moderate technological expense by using an equivalent of two insert modules (an effective sensor thickness of 4 mm). |
|
|
Address |
[Grkovski, Milan; Cindro, Vladimir; Mikuz, Marko; Studen, Andrej; Zontar, Dejan] Jozef Stefan Inst, Ljubljana, Slovenia, Email: milan.grkovski@ijs.si |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Elsevier Science Bv |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0168-9002 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
WOS:000354870700016 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
yes |
|
|
Call Number |
IFIC @ pastor @ |
Serial |
2232 |
|
Permanent link to this record |
|
|
|
|
Author |
Studen, A.; Burdette, D.; Chesi, E.; Cindro, V.; Clinthorne, N.H.; Cochran, E.; Grosicar, B.; Kagan, H.; Lacasta, C.; Linhart, V.; Mikuz, M.; Stankova, V.; Weilhammer, P.; Zontar, D. |
|
|
Title |
Timing performance of the silicon PET insert probe |
Type |
Journal Article |
|
Year |
2010 |
Publication |
Radiation Protection Dosimetry |
Abbreviated Journal |
Radiat. Prot. Dosim. |
|
|
Volume |
139 |
Issue |
1-3 |
Pages |
199-203 |
|
|
Keywords |
|
|
|
Abstract |
Simulation indicates that PET image could be improved by upgrading a conventional ring with a probe placed close to the imaged object. In this paper, timing issues related to a PET probe using high-resistivity silicon as a detector material are addressed. The final probe will consist of several (four to eight) 1-mm thick layers of silicon detectors, segmented into 1 x 1 mm(2) pads, each pad equivalent to an independent p + nn+ diode. A proper matching of events in silicon with events of the external ring can be achieved with a good timing resolution. To estimate the timing performance, measurements were performed on a simplified model probe, consisting of a single 1-mm thick detector with 256 square pads (1.4 mm side), coupled with two VATAGP7s, application-specific integrated circuits. The detector material and electronics are the same that will be used for the final probe. The model was exposed to 511 keV annihilation photons from an Na-22 source, and a scintillator (LYSO)-PMT assembly was used as a timing reference. Results were compared with the simulation, consisting of four parts: (i) GEANT4 implemented realistic tracking of electrons excited by annihilation photon interactions in silicon, (ii) calculation of propagation of secondary ionisation (electron-hole pairs) in the sensor, (iii) estimation of the shape of the current pulse induced on surface electrodes and (iv) simulation of the first electronics stage. A very good agreement between the simulation and the measurements were found. Both indicate reliable performance of the final probe at timing windows down to 20 ns. |
|
|
Address |
[Studen, A.; Cindro, V.; Grosicar, B.; Mikuz, M.; Zontar, D.] Jozef Stefan Inst, SI-1000 Ljubljana, Slovenia, Email: andrej.studen@ijs.si |
|
|
Corporate Author |
|
Thesis |
|
|
|
Publisher |
Oxford Univ Press |
Place of Publication |
|
Editor |
|
|
|
Language |
English |
Summary Language |
|
Original Title |
|
|
|
Series Editor |
|
Series Title |
|
Abbreviated Series Title |
|
|
|
Series Volume |
|
Series Issue |
|
Edition |
|
|
|
ISSN |
0144-8420 |
ISBN |
|
Medium |
|
|
|
Area |
|
Expedition |
|
Conference |
|
|
|
Notes |
ISI:000277738200035 |
Approved |
no |
|
|
Is ISI |
yes |
International Collaboration |
yes |
|
|
Call Number |
IFIC @ elepoucu @ |
Serial |
449 |
|
Permanent link to this record |