PT Journal AU Capra, S Mengoni, D DueƱas, JA John, PR Gadea, A Aliaga, RJ Dormard, JJ Assie, M Pullia, A TI Performance of the new integrated front-end electronics of the TRACE array commissioned with an early silicon detector prototype SO Nuclear Instruments & Methods in Physics Research A JI Nucl. Instrum. Methods Phys. Res. A PY 2019 BP 178 EP 184 VL 935 DI 10.1016/j.nima.2019.05.039 LA English DE ASIC; Charge-sensitive preamplifier; Low-noise applications; Particle spectrometry; Dead time; Silicon detector AB The spectroscopic performances of the new integrated ASIC (Application-Specific Integrated Circuit) preamplifiers for highly segmented silicon detectors have been evaluated with an early silicon detector prototype of the TRacking Array for light Charged Ejectiles (TRACE). The ASICS were mounted on a custom-designed PCB (Printed Circuit Board) and the detector plugged on it. Energy resolution tests, performed on the same detector before and after irradiation, yielded a resolution of 21 keV and 33 keV FWHM respectively. The output signals were acquired with an array of commercial 100-MHz 14-bit digitizers. The preamplifier chip is equipped with an innovative Fast-Reset device that has two functions: it reduces dramatically the dead time of the preamplifier in case of saturation (from milliseconds to microseconds) and extends the spectroscopic dynamic range of the preamplifier by more than one order of magnitude. Other key points of the device are the low noise and the wide bandwidth. ER