PT Journal AU Alvarez, V Herrero-Bosch, V Esteve, R Laing, A Rodriguez, J Querol, M Monrabal, F Toledo, JF Gomez-Cadenas, JJ TI The electronics of the energy plane of the NEXT-White detector SO Nuclear Instruments & Methods in Physics Research A JI Nucl. Instrum. Methods Phys. Res. A PY 2019 BP 68 EP 76 VL 917 DI 10.1016/j.nima.2018.11.126 LA English DE Calorimetry; Front-end electronics; Digital baseline restoration AB This paper describes the electronics of NEXT-White (NEW) detector PMT plane, a high pressure xenon TPC with electroluminescent amplification (HPXe-EL) currently operating at the Laboratorio Subterraneo de Canfranc (LSC) in Huesca, Spain. In NEXT-White the energy of the event is measured by a plane of photomultipliers (PMTs) located behind a transparent cathode. The PMTs are Hamamatsu R11410-10 chosen due to their low radioactivity. The electronics have been designed and implemented to fulfill strict requirements: an overall energy resolution below 1% and a radiopurity budget of 20 mBq unit(-1) in the chain of Bi-214. All the components and materials have been carefully screened to assure a low radioactivity level and at the same time meet the required front-end electronics specifications. In order to reduce low frequency noise effects and enhance detector safety a grounded cathode connection has been used for the PMTs. This implies an AC-coupled readout and baseline variations in the PMT signals. A detailed description of the electronics and a novel approach based on a digital baseline restoration to obtain a linear response and handle AC coupling effects is presented. The final PMT channel design has been characterized with linearity better than 0.4% and noise below 0.4 mV. ER