TY - JOUR AU - Agramunt, J. et al AU - Tain, J. L. AU - Albiol, F. AU - Algora, A. AU - Domingo-Pardo, C. AU - Jordan, M. D. AU - Rubio, B. AU - TarifeƱo-Saldivia, A. AU - Valencia, E. PY - 2016 DA - 2016// TI - Characterization of a neutron-beta counting system with beta-delayed neutron emitters T2 - Nucl. Instrum. Methods Phys. Res. A JO - Nuclear Instruments & Methods in Physics Research A SP - 69 EP - 78 VL - 807 PB - Elsevier Science Bv KW - Beta-delayed neutron emission probability KW - Neutron and beta counters KW - Self-triggered digital data acquisition system KW - Geant4 simulations AB - A new detection system for the measurement of beta-delayed neutron emission probabilities has been characterized using fission products with well known beta-delayed neutron emission properties. The setup consists of BELEN-20, a 4 pi-neutron counter with twenty He-3 proportional tubes arranged inside a large polyethylene neutron moderator, a thin Si detector for beta counting and a self-triggering digital data acquisition system. The use of delayed-neutron precursors with different neutron emission windows allowed the study of the effect of energy dependency on neutron, beta and beta-neutron rates. The observed effect is well reproduced by Monte Carlo simulations. The impact of this dependency on the accuracy of neutron emission probabilities is discussed. A new accurate value of the neutron emission probability for the important delayed-neutron precursor I-137 was obtained, P-n = 7.76(14)%. SN - 0168-9002 UR - https://doi.org/10.1016/j.nima.2015.10.082 DO - 10.1016/j.nima.2015.10.082 LA - English N1 - WOS:000365596200010 ID - Agramunt_etal2016 ER -