TY - JOUR AU - Barrio, J. AU - Etxebeste, A. AU - Lacasta, C. AU - Muñoz, E. AU - Oliver, J. F. AU - Solaz, C. AU - Llosa, G. PY - 2015 DA - 2015// TI - Performance of VATA64HDR16 ASIC for medical physics applications based on continuous crystals and SiPMs T2 - J. Instrum. JO - Journal of Instrumentation SP - P12001 - 12pp VL - 10 PB - Iop Publishing Ltd KW - Solid state detectors KW - Photon detectors for UV KW - visible and IR photons (solid-state) (PIN diodes KW - APDs KW - Si-PMTs KW - G-APDs KW - CCDs KW - EBCCDs KW - EMCCDs etc) KW - Front-end electronics for detector readout KW - Gamma detectors (scintillators KW - CZT KW - HPG KW - HgI etc) AB - Detectors based on Silicon Photomultipliers (SiPMs) coupled to continuous crystals are being tested in medical physics applications due to their potential high resolution and sensitivity. To cope with the high granularity required for a very good spatial resolution, SiPM matrices with a large amount of elements are needed. To be able to read the information coming from each individual channel, dedicated ASICs are employed. The VATA64HDR16 ASIC is a 64-channel, charge-sensitive amplifier that converts the collected charge into a proportional current or voltage signal. A complete assessment of the suitability of that ASIC for medical physics applications based on continuous crystals and SiPMs has been carried out. The input charge range is linear from 20 pC up to 55 pC. The energy resolution obtained at 511 keV is 10% FWHM with a LaBr3 crystal and 16% FWHM with a LYSO crystal. A coincidence timing resolution of 24 ns FWHM is obtained with two LYSO crystals. SN - 1748-0221 UR - https://doi.org/10.1088/1748-0221/10/12/P12001 DO - 10.1088/1748-0221/10/12/P12001 LA - English N1 - WOS:000369998500034 ID - Barrio_etal2015 ER -