TY - JOUR AU - Llosa, G. AU - Barrio, J. AU - Lacasta, C. AU - Callier, S. AU - Raux, L. AU - de La Taille, C. PY - 2011 DA - 2011// TI - First tests in the application of silicon photomultiplier arrays to dose monitoring in hadron therapy T2 - Nucl. Instrum. Methods Phys. Res. A JO - Nuclear Instruments & Methods in Physics Research A SP - S96-S99 VL - 648 PB - Elsevier Science Bv KW - Hadron therapy KW - Compton imaging KW - LaBr3 KW - Continuous crystal KW - SiPM KW - MPPC KW - G-APD AB - A detector head composed of a continuous LaBr3 crystal coupled to a silicon photomultiplier array has been mounted and tested, for its use in a Compton telescope for dose monitoring in hadron therapy. The LaBr3 crystal has 16 mm x 18 mm x 5 mm size, and it is surrounded with reflecting material in five faces. The SiPM array has 16 (4 x 4) elements of 3 mm x 3 mm size. The SPIROC1 ASIC has been employed as readout electronics. The detector shows a linear behavior up to 1275 keV. The energy resolution obtained at 511 keV is 7% FWHM, and it varies as one over the square root of the energy up to the energies tested. The variations among the detector channels are within 12%. A preliminary measurement of the timing resolution gives 7 ns FWHM. The spatial resolution obtained with the center of gravity method is 1.2 mm FWHM. The tests performed confirm the correct functioning of the detector. SN - 0168-9002 UR - https://doi.org/10.1016/j.nima.2010.12.092 DO - 10.1016/j.nima.2010.12.092 LA - English N1 - WOS:000305376900026 ID - Llosa_etal2011 ER -