PT Journal AU R3B Collaboration (Heil, Mea Nacher, E TI A new Time-of-flight detector for the (RB)-B-3 setup SO European Physical Journal A JI Eur. Phys. J. A PY 2022 BP 248 - 19pp VL 58 IS 12 DI 10.1140/epja/s10050-022-00875-8 LA English AB We present the design, prototype developments and test results of the new time-of-flight detector (ToFD) which is part of the R3B experimental setup at GSI and FAIR, Darmstadt, Germany. The ToFD detector is able to detect heavy-ion residues of all charges at relativistic energies with a relative energy precision sigma_Delta E/Delta E of up to 1% and a time precision of up to 14 ps (sigma). Together with an elaborate particle-tracking system, the full identification of relativistic ions from hydrogen up to uranium in mass and nuclear charge is possible. ER