PT Journal AU Granero, D Candela-Juan, C Vijande, J Ballester, F Perez-Calatayud, J Jacob, D Mourtada, F TI Technical Note: Dosimetry of Leipzig and Valencia applicators without the plastic cap SO Medical Physics JI Med. Phys. PY 2016 BP 2087 EP 4pp VL 43 IS 5 DI 10.1118/1.4944784 LA English DE Leipzig applicators; Valencia applicators; skin brachytherapy; Monte Carlo; dosimetry AB Purpose: High dose rate (HDR) brachytherapy for treatment of small skin lesions using the Leipzig and Valencia applicators is a widely used technique. These applicators are equipped with an attachable plastic cap to be placed during fraction delivery to ensure electronic equilibrium and to prevent secondary electrons from reaching the skin surface. The purpose of this study is to report on the dosimetric impact of the cap being absent during HDR fraction delivery, which has not been explored previously in the literature. Methods: GEANT4 Monte Carlo simulations (version 10.0) have been performed for the Leipzig and Valencia applicators with and without the plastic cap. In order to validate the Monte Carlo simulations, experimental measurements using radiochromic films have been done. Results: Dose absorbed within 1 mm of the skin surface increases by a factor of 1500% for the Leipzig applicators and of 180% for the Valencia applicators. Deeper than 1 mm, the overdosage flattens up to a 10% increase. Conclusions: Differences of treating with or without the plastic cap are significant. Users must check always that the plastic cap is in place before any treatment in order to avoid overdosage of the skin. Prior to skin HDR fraction delivery, the timeout checklist should include verification of the cap placement. (C) 2016 American Association of Physicists in Medicine. ER