PT Journal AU Llosa, G Barrillon, P Barrio, J Bisogni, MG Cabello, J Del Guerra, A Etxebeste, A Gillam, JE Lacasta, C Oliver, JF Rafecas, M Solaz, C Stankova, V de La Taille, C TI High performance detector head for PET and PET/MR with continuous crystals and SiPMs SO Nuclear Instruments & Methods in Physics Research A JI Nucl. Instrum. Methods Phys. Res. A PY 2013 BP 3 EP 5 VL 702 DI 10.1016/j.nima.2012.08.099 LA English DE Monolithic crystals; SiPM; MG-APD; PET; High resolution; Position determination AB A high resolution PET detector head for small animal PET applications has been developed. The detector is composed of a 12 mm x 12 mm continuous LYSO crystal coupled to a 64-channel monolithic SiPM matrix from FBK-irst. Crystal thicknesses of 5 mm and 10 mm have been tested, both yielding an intrinsic spatial resolution around 0.7 mm FWHM with a position determination algorithm that can also provide depth-of-interaction information. The detectors have been tested in a rotating system that makes it possible to acquire tomographic data and reconstruct images of Na-22 sources. An image reconstruction method specifically adapted for continuous crystals has been employed. The Full Width at Half Maximum measured from a point source reconstructed with ML-EM was 0.7 mm with the 5 mm crystal and 0.8 mm with the 10 mm crystal. ER