%0 Journal Article %T Description and commissioning of NEXT-MM prototype: first results from operation in a Xenon-Trimethylamine gas mixture %A NEXT Collaboration (Alvarez, V. et al %A Carcel, S. %A Cervera-Villanueva, A. %A Diaz, J. %A Ferrario, P. %A Gil, A. %A Gomez-Cadenas, J. J. %A Laing, A. %A Liubarsky, I. %A Lorca, D. %A Martin-Albo, J. %A Martinez, A. %A Monrabal, F. %A Muñoz Vidal, J. %A Nebot-Guinot, M. %A Rodriguez, J. %A Serra, L. %A Simon, A. %A Sorel, M. %A Yahlali, N. %J Journal of Instrumentation %D 2014 %V 9 %I Iop Publishing Ltd %@ 1748-0221 %G English %F NEXTCollaborationAlvarez_etal2014 %O WOS:000336123200070 %O exported from refbase (https://references.ific.uv.es/refbase/show.php?record=1800), last updated on Wed, 17 Jul 2019 09:56:38 +0000 %X A technical description of NEXT-MM and its commissioning and first performance is reported. Having an active volume of similar to 35 cm drift x 28 cm diameter, it constitutes the largest Micromegas-read TPC operated in Xenon ever constructed, made by a sectorial arrangement of the 4 largest single wafers manufactured with the Microbulk technique to date. It is equipped with a suitably pixelized readout and with a sufficiently large sensitive volume (similar to 23 l) so as to contain long (similar to 20 cm) electron tracks. First results obtained at 1 bar for Xenon and Trymethylamine (Xe-(2%) TMA) mixture are presented. The TPC can accurately reconstruct extended background tracks. An encouraging full-width half-maximum of 11.6% was obtained for similar to 29 keV gammas without resorting to any data post-processing. %K Time projection chambers %K Particle tracking detectors (Gaseous detectors) %K Double-beta decay detectors %R 10.1088/1748-0221/9/03/P03010 %U http://arxiv.org/abs/1311.3242 %U https://doi.org/10.1088/1748-0221/9/03/P03010 %P P03010 - 22pp